The alignment error of Elinox ELS-7500EX E-beam Lithography system is examined, as a project of Graduate Student Fellow Program. This report shows the alignment error of 60 and 30 nm in the x- and y-direction, respectively, confirming the tool specification of 60 nm alignment error.